Specs

 

Standards:
Hearing Aid Analyzer As Per ANSI S3.22-2009, IEC 60118-7 2005

Speaker Output:
Max 95 dB SPL

Frequency Range:
125 Hz - 8000 Hz ± 0.5%

Coupler Mic:
Max Input Level: 140 dB SPL

Reference Mic:
Omni Directional Microphone

Battery Simulator Output:
Adjustable Output Voltage: 0.1V - 1.5V In 100mV Steps Accuracy ± 5%. Current Measurement: 20uA - 20mA Accuracy ± 5%. Battery Pills Provided: 10A, 13, 312 And 675

Magnetic Loop:
31.6mA/m Magnetic Strength,
per ANSI Standard

Communication Port: USB

Power Requirements:
USB Power

Dimensions:
Approx. 24 cm x 25 cm x 16 cm
(L x W x H)

Net Weight:
Approx. 1,8 kg / Approx 7 lbs

Enclosure:
Acoustically dampened enclosure

Environmental Requirements:
Working temperature range
from 10°C to 35°C

HIT plus

AVANT HIT+

Hearing Instrument Test Chamber


HIT plus openThe AVANT HIT+ is an advanced programmable PC-based hearing instrument test chamber. AVANT HIT+ offers comprehensive ANSI and IEC test protocols that offer user- selectable automated testing.

This device is USB-powered, eliminating direct power connections to the wall. With its small footprint (24 x 25 x 16 cm), AVANT HIT+ is discreet and space-saving. All test results are seamlessly stored within the NOAH™ patient database and it is also TIMS® and Sycle.net compatible.

The system provides all standard couplers, battery pills, cables and software as part of the standard package.

AVANT HIT+ Data Sheet
HIT Box Coupler Assembly


Software Update  |  Installation Manual Rev 5 User Manual Rev 4